The COMPACS® system
Real-time vibration diagnostics
of machinery with probability close to 1
Русский Русский     EnglishEnglish 
Menu
Home
Products
Customers & responses
Personnel training
About company
News
Publications
Contacts
Awards
  • 2016 "Russia’s Honoured Engineer"
  • 2016 "Recognition"
  • 2016 "Import Substitution"
  • 2016 "Innovations and Quality"
  • 2015 "Outstanding manager"
  • 2015 "100 best goods of Russia"
  • 2015 "ESQR’s Quality Achievements Awards"
  • 2014 "Best Quality of Rolling Stock and Complex Technical Systems Contest conducted by Russian Railways"
  • 2014 "Highly-efficient  company"
  • 2014 "Reliable supplier"
  • 2014 "Leader of the Industry"
  • 2014 "Accountant of the Year"
  • 2014 "Technological breakthrough"
  • 2013 "Business elite of Russia"
  • 2013 "Top 100 of Russian Goods"
Tags Cloud
Follow us
Facebook Twitter LinkedIn YouTube Google+ RSS
Certification
In 2001 the quality management system of SPC Dynamics was put through facultative certification. In 2016 it was recertified for compliance with standards ISO 9001:2008. Its results confirmed a high level of goods and services quality management. In 2001 the quality management system of SPC Dynamics was put through facultative certification. In 2016 it was recertified for compliance with standard GOST ISO 9001-2011. Its results confirmed a high level of goods and services quality management.
Orphus system
Home Products Calibrator 8003

Calibration of velocity sensors, vibration accelerometers and thermocouples: calibrator 8003

Print
Calibrator 8003

Calibrator 8003

Calibrator 8003 is designed for verification and calibration of systems and devices, measuring parameters of vibration, voltage and direct and alternating current forces, as well as temperature by means of a THK(L) type thermocouple.

Since the calibrator can operate in off-line mode (due to built-in rechargeable batteries), it provides calibration tests of multi-channel measuring systems directly on the site without dismantling.

Calibrator 8003 allows to reproduce mechanical oscillations, voltage and direct and alternating current forces within frequency and amplitude ranges, a thermopower in a THK(L) type thermocouple, as well as an active resistance.

The calibrator can be supported by the COMPACS® software providing the system measuring channels calibration. On the basis of calibrator 8003 and the COMPACS® vibration monitoring system, there have been developed methods allowing to reduce considerably duration and costs for system measuring channels calibration.

Calibrator 8003 advantages

  • power supply both from alternating current mains and built-in rechargeable batteries;
  • manual and automatic operation modes, allowing the system functioning in one of the programs, set by the user.

Calibrator 8003 structure

  • setting device for electric signals generation;
  • vibration table, performing a sine electric signal conversion into mechanical oscillations of a vibration table.

The calibrator is registered in the State Register of Measuring Equipment under No 25732-03 (Certificate of instrumentation type approval RU.C.28.004.А No 16074).

Warranty for supplied equipment - 12 months.

Basic characteristics
RMS values range for:
vibration acceleration, m/s2
velocity, mm/s
vibration displacement, μm
Frequency range for**:
vibration acceleration, Hz
velocity, Hz
vibration displacement, Hz
Voltage range for:
direct current, mV
alternating current (RMS), mV
Frequency range for:
voltage, Hz
current force, Hz
Range for direct current force, mА
RMS range for alternating current, mА
Active resistance range, Ohm
Range for a unique sensor curve of a THK(L) thermocouple, °С
Measuring range of a conversion ratio on the calibrated
VMT (vibration measuring transducer) charge, pC/m-2
Maximal load on vibration table, g
Acceleration harmonics factor, %, not exceeding
within frequency range from 10 to 20 Hz
within frequency range from 20 to 3000 Hz
Relative ratio of transverse components in mechanical
oscillations, %, not exceeding
Operation temperature range, °С
Time of continuous operation in a mode of vibration acceleration
10 m/sec2 at F=159,2 Hz at ml=200g, hours, not less
Calibrator weight, kg

1...100*
1...100
1...1000

10...3000
10...1000
10...200

0,5...2000
1...1000

3...9999
3...1000
0,5...100
1...50
10...511
-50...+800

2...100 
200

10
5

20
-10...+40

4 
18,2
Technical characteristics
Relative basic error, %:
vibration acceleration within range (10...20) Hz
vibration acceleration within range (20...3000) Hz
velocity within range (10...20) Hz
velocity within range (20...1000) Hz
vibration displacement within range (10...20) Hz
vibration displacement within range (20...200) Hz

±3,0%
±2,0%
±3,0%
±2,0%
±4,0%
±2,5%
direct current voltage within ranges, %:
±(0,5...4,99) мВ
±(5,0...99,9) мВ
±(100...2000) мВ
alternating current voltage:
within frequency range of (3,0...10) Hz
within RMS range (1,0...4,99) mV
within RMS ranges (5,0...99,9) mV and (100...1000) mV
within frequency range (10...9999) Hz4
within RMS range (1,0...4,99) mV
within RMS ranges (5,0...99,9) mV and (100...1000) mV
accuracy class:
0,4/0,25
0,1/0,01
0,1
accuracy class:

6,0/2,5
0,6/0,15

3,0/2,5
0,6/0,15
direct current force:
±(0,5...4,99) mA
±(5,0...100) mA
alternating current force:
within frequency range (3,0...10) Hz
within RMS range (1,0...4,99) mА
within RMS range (5,0...50) mА
within frequency range (10...1000) Hz
within RMS range (1,0...4,99) mА
within RMS range (5,0...50) mА
accuracy class:
1,6/0,15
1,6
accuracy class:

6,0/2,5
4,0/0,3

2,5/2,0
1,6/0,2
Absolute basic error oa unique sensor curve
of a THK(L) type thermocouple, °С
-40...+100
(-50...-40) and (+100...+800)


±0,3
±1,0
Absolute basic error of resistance, Ohm ±2,0
Relative basic error of a conversion ratio measuring on
the calibrated VMT (vibration measuring transducer) charge, %

±2,0


      * with load mass ml= 40 g

      ** with a normalizable error. With an external vibration-measuring transducer
      the frequency range of vibration acceleration is from 10 to 9999 Hz

      Specifications can be changed without prior notification